29 Sept 2021 · Coroner Kevin McLoughlin · West Yorkshire (East)

The coroner's concerns

  1. 1. loose connection of Philips Respironics AF 541 mask to BIPAP ventilator

    “The Philips Respironics AF 541 mask connects to the tubing, linking it to the BIPAP ventilator by means of a 'push on' connection (rather than a fitting involving positive engagement). Evidence taken at the Inquest indicates this connection has come undone on other occasions as well. The introduction of a filter at the site of this union increases the potential for the joint to come apart. Consideration should be given to installing a more robust docking mechanism which is less vulnerable to working”