Ben Harrison
10 May 2024 · Coroner Kate Robertson · North Wales (East and Central)
The coroner's concerns
1. confusion of two valves on oxygen cylinder
“Evidence was heard during the Inquest that a CD Oxygen cylinder manufactured by BOC was used during the resuscitation of [the deceased]. I was shown during the Inquest how the oxygen cylinder is operated.”
2. lack of response to safety concerns
“It is understood that BCUHB have referred numerous concerns to BOC over recent years and some minor amendments have been made to the cylinders. BCUHB also reported to the Medicines and Healthcare products Regulatory Agency (MHRA) on 6 October 2022 under The Yellow Card Scheme. No response was formally received.”
3. unsafe operation of CD Oxygen cylinders
“I remain concerned that the CD Oxygen cylinders manufactured by BOC which, it is understood, supply most if not all Health Boards in Wales (under procurement processes) are unsafe for use in heightened / pressurised situations in that it is not overtly clear how the cylinders are to be operated with the confusion of the two valves. This is evidenced by very similar ongoing issues identified by BCUHB even with adequate training of staff.”